Package: Software on CD, hardware A/D converter (internal PCI, external USB or external TCP/IP for network), hardware key, detector and trigger cables, manual and mousemat.
System configuration: Any 32-bit Windows system (98, Me, NT, 2000, XP)
Recommended configuration for Win 2000/ XP: PC Pentium/900 MHz, 256 MB RAM, monitor 1024x768, 64K colors (16 bit High Color), CD ROM, free PCI slot/USB connector for A/D converter.
Minimal configuration for Win 98: PC Pentium/200 MHz, 32 MB RAM, monitor 800x600, 256 colors, CD ROM, free PCI slot/USB connector for A/D converter.
Number of instruments/channels: Up to 4 chromatographs (instruments) at a time, each up to 4 detectors (channels). 1 chromatograph (instrument) - 1 time base Supported chromatographs: Any detector with voltage output, up to 30m distance Measuring ranges: Bipolar: 156, 1 250, 10 000 mV, integration frequency up to 100 Hz, 24-bit resolution GLP/ 21 CFR Part11 requirements:
Password protection
Password expiration
Electronic signature
Audit trails
User account with access rights
Interface with auto samplers: The station interfaces with all auto samplers (as a slave or master), direct control available (refer to the list of controlled instruments) of some auto samplers.
Integration: Separate integration table for every detector, 27 integration parameters (peak width, threshold, tangent slope ratio etc.), integration parameters programmable in time, automated reintegration.
Calculation types: Both without and with calibration (internal and external standard method), parameters of individual peaks for assessing efficiency of the column and the chromatographic system as a whole, SST Module for establishing deviations and reproducibility of some selected parameters.
Calibrations: 6 types of calibration curves, up to 20 levels, reference peaks, groups, unlimited number of standards (peaks) Work with chromatograms: Overlay, mathematical operations with chromatograms, custom labels and settings for chromatograms.
Automation: Sequences, automatic launch of selected commands or applications immediately following chromatogram acquisition – Post Run, Batch.
Presentation of results: Integrated customisable table of results, columns with user-defined calculation, summary table, and export in text or database format.
Custom calculations: 12 predefined mathematical operators, 15 basic and 4 summary functions.
Protocol: Custom protocols, print preview option Data export: ASCII, AIA, dBase Data import: CSW17, CSW32, ASCII, AIA.
Multi-user environment: Selectable system of user accounts with independently customisable parameters and appearance for individual users.
Network environment: Easy off-line (at the file level) data sharing among all stations in a local network Add on modules.
Extensions: Instrument and auto sampler control modules, GPC software, Test IQ, Validation Kit, SST Module, Clarity Evaluation software (off-line version for reprocessing on another PC), LIMS Interface,
Support, warranty, Automatic software update from Internet, Free software updates, Free access to knowledge database, Online technical support, 3 years hardware warranty,
|